Epsilon

UEi KM945

$1,560.00

Brand: UEi
Part #KM945
*UEi KM940 is Discontinued and replaced with KM945*

UEi KM945 Industrial Analyzer
The KM945 handheld analyzer is ideal for industrial boiler flue gas measurement and analysis. It operates on all boiler types and can display 8 parameters simultaneously showing measurements and calculations of boiler efficiency and pollution.

Category: Tag:

The KM945 base unit measures O2, flue and net temperature and calculates CO2, boiler efficiency, excess air and losses. Specify the toxic gas sensors to meet your needs.

Features:

  • Measures O2
  • Calculates CO2, efficiency and excess air
  • Large 4 line display
  • Long life rechargeable battery
  • 150 memory positions


SPECIFICATIONS

Function Resolution Accuracy Range
Measurement Flue Measurement with Probe 1.0°F/C ±2.0°F/C, ±0.3% of rdg 32° – 1112°F, (0° – 600°C)
Inlet Temperature 0.1°F/C ±1.0°F/C, ±0.3% of rdg 32° – 122°F, (0° – 50°C)
Pressure 0.1 mbar ±2% of full scale +150 mbar to – 150 mbar
Gas Measurement* Oxygen (02) 0.1% ±0.2% 0 – 21%
Carbon Monoxide
standard: H compensated
(optional)
1 ppm ± 20 ppm <400 ppm
±5% >5000 ppm
± 10% <5000 ppm
0 – 10,000 ppm
Carbon Monoxide
High range
(optional)
0.01% ±5% reading from 0.1% to 10% 0 ~ 10%
Nitric Oxide (NO)
standard
(optional)
1 ppm ±5 ppm <100 ppm
±5%>100 ppm
0 – 5000 ppm
Nitric Oxide (NO)
low range
(optional)
1 ppm ±2 ppm <30 ppm
±5 ppm <30 ppm
0 – 100 ppm
Nitrogen Dioxide 1 ppm ±5 ppm<100 ppm
±10 ppm<500 ppm
±5%>500 ppm
0 – 1000 ppm
Sulphur Dioxide 1 ppm ±5 ppm<100 ppm
±5%>100 ppm
0 – 5000 ppm
Calculations Carbon Dioxide 0.1% ±0.3% reading 0 – 99.9%
Losses 0.1% ±1.0% reading 0 – 99.9%
Efficiency 0.1% ±1.0% reading 0 – 99.9%
Excess Air 0.1% ±0.2% 0 ~ 2885.0%
Temp (Nett) 1.0°C/F ±2°C ±0.3% reading 32 ~ 1112°F (0 ~ 600°C)
CO/CO2 Ratio 0.0001 ±0.0001 0 ~ 0.9999
Poison Index 0.01% ±0.01 0 ~ 99.99
Preprogrammed Fuels Natural Gas, Town Gas, Gascor, Light Oil, Heavy Oil, Propane, Butane, Anthrocite, Coke, Coal, Kinsale Gas
Dimensions Weight 1 Kg
Handset 220 x 55 x 120 mm
Probe (L) 420 x (Dia) 8mm with stainless steel shaft, K-type thermocouple and 3m hose
Ambient operating range 0°C to 45°C/10% to 90% RH non condensing
Power supply (battery charger) Input: 110V AC / 220V AC nominal, Output: 12V AC off load
Battery life >6 hours from full charge

* Toxic gas sensors need to be specified at time of purchase or may be added at a later time.